Seitenansicht 11
̶
9
̶
3-5) Range/ Filter Function
Range Configuration
Range configuration varies depending on each connected Sensor.
Refer to "12. Specification" (P.45) in this manual.
Auto-ranging
Auto-ranging is available only in Normal Recording mode. It isn't
available on a Sensor providing one single range.
(e.g. Voltage Sensor "KEW8309" for KEW5020)
Range Hold
Select the ranges with
mark to fix measurement range under
Normal Recording mode.
Measurement range in Trigger/ Capture Recording mode is decided
depending on a detection level. Switching of measurement ranges
with the button isn't possible.
Filter Function
Lowpass Filter works when Filter Function is ON
mark lights up
and filter out the frequency in higher harmonics band.
(Cutoff Frequency: approx 160Hz, attenuation rate -24dB)
Range/ Filter Function for each Recording mode
Recording mode Function
Normal recording
* Switching ranges, On/Off Filter
* Auto-ranging is available.
Trigger/ Capture
recording
* On/Off Filter
* Range corresponds to detection level
* Auto-ranging isn't available.
Power Quality Analysis
[KEW5020 only]
* On/Off Filter
* Auto-ranging isn't available.